Source[1|2]]:phase:initiate (no query form), Source[1|2]]:pm[:deviation – Tektronix AFG3000 User Manual

Page 135

Advertising
background image

Syntax and Commands

AFG3000 Series Arbitrary/Function Generators Reference Manual

3-61

[SOURce[1|2]]:PHASe:INITiate (No Query Form)

This command synchronizes the phase of CH 1 and CH 2 output waveforms. The
arbitrary/function generator performs the same operation if you specify either
SOURce1 or SOURce2.

If your arbitrary/function generator is single-channel model, this command is not
supported.

Group

Source

Syntax

[SOURce[1|2]]:PHASe:INITiate

Arguments

None

Examples

SOURce1:PHASe:INITiate
synchronizes the phase of CH 1 and CH 2 output signals.

[SOURce[1|2]]:PM[:DEViation]

This command sets or queries the peak frequency deviation of PM modulation for
the specified channel.

Group

Source

Syntax

[SOURce[1|2]]:PM:DEViation {<deviation>|MINimum|MAXimum}

[SOURce[1|2]]:PM:DEViation? {MINimum|MAXimum}

Arguments

<deviation>::=<NR3>[<units>]
where
<NR3> is the phase deviation.
<units>::=[RAD | DEG]

If <units> are omitted, RAD is specified automatically. The setting ranges are:

RAD

–1 PI to +1 PI, relative to phase value

DEG

–180 to +180, in 1 degree steps, relative to phase value

Returns

<deviation>

Examples

SOURce1:PM:DEViation MAXimum
sets the maximum value for the CH 1 phase deviation.

Advertising