AFG3000 Series Arbitrary/Function Generators Reference Manual

3-17

CALibration[:ALL]

The CALibration[:ALL] command performs an internal calibration.

The CALibration[:ALL]? command performs an internal calibration and returns
0 (Pass) or a calibration error code.

Group

Calibration and Diagnostic

Related Commands

*CAL?

Syntax

CALibration[:ALL]

CALibration[:ALL]?

Arguments

None

Returns

<NR1>

Examples

CALibration:ALL
performs an internal calibration.

CALibration:ALL?
performs an internal calibration and returns results. For example, it might return 0,
which indicates that the calibration completed without any errors.

*CLS (No Query Form)

This command clears all the event registers and queues, which are used in the arbi-
trary/function generator status and event reporting system.

Group

Status

Syntax

*CLS

Arguments

None

Examples

*CLS
clears all the event registers and queues.

NOTE.

The self-calibration can take several minutes to complete. During this time,

the arbitrary/function generator does not execute any commands. Do not power off
the instrument during the self-calibration.