VEGA VEGASON 71 - 1 … 75 - 1 User Manual
Page 19
Adjustment
71 - 1 … 75 - 1
19
Selection of adjustment procedure
The type of empty adjustment (mode 11) and full
adjustment (mode 12) is determined with the para-
meter adjustment of this mode.
The type of adjustment depends on the local condi-
tions.
Mode description
Mode 10 = 1
means adjustment in m without product level change, i.e.
- with the adjustment procedure the distances in m determined from the vessel drawings
are programmed relating to the levels 0 % and 100 %
- level is not important.
Mode 10 = 3
means adjustment in % with product level change, i.e.
- with this adjustment procedure the levels are programmed in % and the respective
measured distance (sensor … level) is calculated automatically
- product level change is needed.
Programming example
Change of mode 10 = 3 of mode 10 = 1 ..................
• select mode 0 - 10 in MODEFIELD
• program figure 1 in PARAMETER FIELD
• then store with STO-button
Demonstration
Mode 10 = 1
Mode 10 = 3
Min. distance
^ 100 %
partly filling
= e.g. 75 %
partly emptying
= e.g. 25 %
Max. distance
^ 0 %
MODE 0 - 10