VEGA VEGASON 71 - 1 … 75 - 1 User Manual

Page 25

Advertising
background image

Linearization

71 - 1 … 75 - 1

25

Programming example:

Index marker 01 ........................................................

• index marker 01

4.H.01

select level percentage
MODEFIELD

Level percentage ......................................................

• acc. to linearity protocol program 12.0 % in

PARAMETER FIELD

• then store with STO-button

• index marker 01

4.L.01

select volume percentage
MODEFIELD

Volume percent ........................................................

• acc. to linearity protocol program 5 % in

PARAMETER FIELD

• then store with STO-button

Index marker 02 .......................................................

• index marker 02

4.H.02

select and program the data as
described above etc.

After programming is finished ..................................

i.e. transmission of all data noted in the linearity
protocol,
-

reset factory settings

-

mode 0 - 15 = 3 (volume percentage)

-

activate linearization curve 4 in mode 0 - 18
(mode 0 - 18 = 4)

Same procedure for linearization curve 5 and 6.