Linearization
71 - 1 … 75 - 1
25
Programming example:
Index marker 01 ........................................................
• index marker 01
4.H.01
select level percentage
MODEFIELD
Level percentage ......................................................
• acc. to linearity protocol program 12.0 % in
PARAMETER FIELD
• then store with STO-button
• index marker 01
4.L.01
select volume percentage
MODEFIELD
Volume percent ........................................................
• acc. to linearity protocol program 5 % in
PARAMETER FIELD
• then store with STO-button
Index marker 02 .......................................................
• index marker 02
4.H.02
select and program the data as
described above etc.
After programming is finished ..................................
i.e. transmission of all data noted in the linearity
protocol,
-
reset factory settings
-
mode 0 - 15 = 3 (volume percentage)
-
activate linearization curve 4 in mode 0 - 18
(mode 0 - 18 = 4)
Same procedure for linearization curve 5 and 6.