VEGA VEGASON 71 - 1 … 75 - 1 User Manual

Page 41

Advertising
background image

Limitation of the operating range

71 - 1 … 75 - 1

41

Operating range

MODE 1 - 05 … 1 - 08

The operating range can be limited in this mode by
programming of
-

start of operating range and

-

end of operating range in case of failure.

Echoes outside this limitation are ignored, e.g.
-

false echoes in the narrow range

-

multiple echoes by reflection in empty vessels
(see programming example)

Note:
In case of multiple reflections, a limited operating
range may not be sufficient. The end of operating
range should be programmed min. 10 % > than the
max. vessel height.

min. distance (e.g. VEGASON 74 - 1)

max. vessel distance

end of operating range

max. distance (e.g. VEGASON 74 - 1)

Programming example:

Indication of distance in m ........................................

• can be indicated in mode 1 - 05

Indication of gain in dB .............................................

• can be indicated in mode 1 - 06

Start of operating range ...........................................

• mode 1 - 07, is here indicated as programming

example

End of operating range .............................................

• select mode 1 - 08 in MODEFIELD
• program 14.00 m in PARAMETER FIELD
• then store with STO-button

0 m

Mode 07 / 08

12

14

0,8

20

false echo by
reflections

Advertising