VEGA VEGASON 66 Foundation Fieldbus User Manual

Page 12

Advertising
background image

The reference plane for the measuring range (version A) is the

lower edge of the flange. The reference plane for versions B, C

and D is the lower edge of the transducer.
Make sure that for all versions a min. distance - the so called

dead zone - below the reference plane is maintained in which a

measurement is not possible. The exact value of the dead

zone is stated in the "Technical data" in the "Supplement".

Information:

If the medium reaches the transducer, buildup can form on it

and cause faulty measurements later on.

1

3

2

100%

0%

4

Fig. 3: Version A - Measuring range (operating range) and max. measuring

distance

1

full (dead zone)

2

empty (max. measuring distance)

3

Measuring range

4

Reference plane

Measuring range

12

VEGASON 66 - Foundation Fieldbus

Mounting

28795

-EN

-060731

Advertising