7 Rotate the connection head in such a way that the cable

entry points downwards (moisture can drain off). Fasten

the cylinder with screws (4); torque max. 5 Nm

8 Insert the plug at the end of the transducer cable into the

socket on the remote electronics

1

2

3

4

Fig. 7: Mounting - Version D

1

Groove for locking the connection head

2

Marking notch

3

Hexagon nut SW 46

4

Terminal screws, connection head (hexagon spanner - size 4)

4.3 Mounting instructions

Mount VEGASON 66 at least 500 mm (19.7 in) from the vessel

wall. If the sensor is installed in the center of vessels with

dished or round tops, multiple echoes can arise. These can,

however, be suppressed by an appropriate adjustment (see

"Setup").
If this distance cannot be maintained, a false echo storage

should be carried out during setup. This applies particularly if

buildup on the vessel wall is expected. In such case, we

recommend repeating the false echo storage later on with

existing buildup.

Installation location

18

VEGASON 66 - Foundation Fieldbus

Mounting

28795

-EN

-060731