4 instrument repair – VEGA VEGASON 66 Foundation Fieldbus User Manual

Page 42

Advertising
background image

In Ex applications, only an instrument and an electronics

module with appropriate Ex approval may be used.

If there is no electronics module available on site, one can be

ordered from the VEGA agency serving you.
The order data of the sensor must be downloaded into the new

electronics module. This can be done:

l

at the factory by VEGA

l

or on site by the user

In both cases, the sensor serial number is necessary. The

serial numbers are stated on the type label of the instrument or

on the delivery note.

Information:

When loading on site, the order data must be downloaded from

the Internet (see Operating Instructions manual "Oscillator").

The oscillators are adapted to the respective sensor and differ

in their signal output or in their power supply. You can find a

suitable oscillator in the following overview.
Electronics module SN-E.64F suitable for VEGASON 64, 65,

66 - Foundation Fieldbus:

l

SN-E.64FX (X = without approvals)

l

SN-E.64FGX (GX = approval GX acc. to VEGA product

list).

8.4 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) in the Internet from

our homepage www.vega.com under: "Downloads - Forms

and Certificates - Repair form".
By doing this you help us carry out the repair quickly and

without having to call back for needed information.

l

Print and fill out one form per instrument

l

Clean the instrument and pack it damage-proof

l

Attach the completed form and possibly also a safety data

sheet to the instrument

l

Please contact the agency serving you for the address of

the return shipment

Sensor serial number

Assignment

Foundation Fieldbus

Maintenance and fault rectification