2

3

1

4

~

=

Power supply

VEGACONNECT 3

PACTware /

TM

Fig. 23: Connecting the PC via HART to the signal cable
1

RS232 connection

2

VEGAFLEX 66

3

HART adapter cable for VEGACONNECT 3

4

HART resistor 250 Ω

N

ecessary components:

l

VEGAFLEX

66

l

PC

with PACTware and suitable VEGA DTM

l

VEGACONNECT

3

with HART adapter cable (art. no.

2.25397)

l

HART

resistor approx. 250 Ω

l

P

ower supply unit

Note:
W

ith power supply units with integrated HART resistance

(

internal resistance approx. 250 Ω), an additional external

resistance is not necessary. This applies, e. g. to the VEGA
instruments VEGATRENN 149A, VEGADIS 371, VEGAMET
381). A

lso usual Ex separators are most of the time equipped

with a sufficient current limitation resistor. In such cases,
VEGACONNECT

3

can be connected parallel to the 4 … 20 mA

cable.

7.2 Parameter adjustment with PACTware

F

urther setup steps are described in the operating instructions

manual "DTM Collection/PACTware" attached to each CD and
which can also be downloaded from our homepage. A detailed
description is available in the online help of PACTware and the
VEGA DTM

s.

Connection via HART

40

VEGAFLEX

66 • 4 … 20

mA/HART

7 S

etup with PACTware and other adjustment programs

32762

-EN

-080716