VEGA VEGAFLEX 66 (-110…+400°C) 4 … 20 mA_HART User Manual

Page 56

Advertising
background image

P

ermissible residual ripple

-

<

100 H

z

U

ss

<

1 V

-

100 H

z … 10 kHz

U

ss

<

10

mV

L

oad

see diagram

1000

750

500

250

14

18

16

20

22

24

26

28

30

32

34

36

V

4

1

2

3

Fig. 30: Voltage diagram
1

HART load

2

Voltage limit EEx-ia instrument

3

Voltage limit non-Ex/Exd instrument

4

Supply voltage

Electrical protective measures

P

rotection, depending on housing version

-

P

lastic housing

IP

66/IP 67

-

A

luminium housing, stainless steel

housing - investment casting, stain-
less steel housing - electro-polished

IP

66/IP 68 (0.2

bar)

4)

-

A

luminium and stainless housing,

investment casting (optionally avail-
able)

IP

66/IP 68 (1

bar)

O

vervoltage category

III

P

rotection class

II

Functional safety (SIL)

F

unctional safety according to IEC 61508/IEC 61511

-

S

ingle channel architecture (1oo1D)

up to SIL2

-

double channel diversitary redun-
dant architecture (1oo2D)

up to SIL3

4)

A

suitable cable is the prerequisite for maintaining the protection class.

56

VEGAFLEX

66 • 4 … 20

mA/HART

10 S

upplement

32762

-EN