Allen key size 2.5 (optional)

l

Flat-headed screwdriver size 1.5 x 10

l

Flat-headed screwdriver size 4 mm

l

Side cutters

l

Gripper tool or manipulator for source capsule

l

Possibly extension rod M6

l

Remove padlock (1)

l

Loosen locking screw (2) with a flat head screwdriver 1.5 x 10

l

Remove the seal of the capstan screw (3) with the side cutting
pliers and pull out the seal wire

l

Unscrew the capstan screw (3) until the fixing strap (4) can be
folded up

l

Fold up the fixing strap (4)

Fig. 13: VEGASOURCE - version D in extraction position

l

If necessary, screw an extension rod (9) with thread M6 into the
thread of the source holder (5). With this extension you can
increase the distance to the emitter

l

Unscrew source holder (5) with spanner (size 13) and pull it out of
the source container

l

Only when exchanging the emitted: Insert the blade of a
screwdriver (4 mm) into the slot (6) on the source holder (5) and
press out the source capsule (7) or unscrew the source capsule
with a screwdriver (2 mm). Let the source capsule (7) fall into a
shielded vessel

Remove source holder

Insert or exchange emit-
ter

Emitter loading and exchange • VEGASOURCE 31 35

15

2 Loading or emitter exchange

39155

-