VEGA VEGASOURCE 35 Emitter loading and exchange User Manual

Page 7

Advertising
background image

l

Remove the seal between holding pin (4) and capstan screw (5)
with side cutting pliers and pull out the seal wire

l

Unscrew the capstan screw (5) with a flat headed screwdriver,
size 1.5 x 10

l

Push in the holding pin (4) and move the source holder (9) over the
pin until the arrow mark points to the holding pin (4)

l

Pull out the source holder (9)

l

Place the cover on the source container to protect against pollution
until reinstallation of the source holder

l

If available, remove safety bolt (6) Allen key size 2.5

l

Swivel the cover (7) in front on the thin cylinder of the source
holder (9) to the side. To do this, lift the cover (7) slightly with a
small slotted screwdriver

l

Only when exchanging the emitter: Allow the source capsule (8) to
fall out downwards into a shielded vessel

l

Insert new source capsule (8) and close cover (7). For insertion
use a suitable gripping device or a manipulator

l

Screw in the safety bolt (6) (optional)

l

Completely insert the source holder (9) into the source container
and turn it until the holding pin (4) pops out

l

Continue turning the source holder (9) up to position OFF

l

Insert the key lock (2) with key and lock. Detach the key

l

If necessary, remove extension rod (1)

l

Screw in capstan screw (5) and secure together with the retaining
pin (4) with new seal

l

Remove existing type labels (when exchanging the emitter)

l

Unpack the type label set and letter it respectively. You can find
further information in chapter "Safety instructions - Labelling"

1

1

3

2

2

4

Fig. 5: Fasten type label
1

Type label - source container (already on the source container)

2

Type label - Source (107 x 17 mm)

3

Adhesive type label - Source (48 x 30 mm)

4

Adhesive label "Radioactive"

l

Letter the adhesive type label (48 x 30 mm)n. Clean the surface
provided on the source holder and attach the adhesive type label

l

Letter the type label (107 x 17 mm) and fasten it with two grooved
pins to the neck of the source container

Insert or exchange emit-
ter

Insert source holder

Attach or replace the
type labels

Emitter loading and exchange • VEGASOURCE 31 35

7

2 Loading or emitter exchange

39155

-

EN

-110210

Advertising