Insert the source holder (6) completely into the source container
l
Turn the source holder until the arrow mark (A) is above the OFF
mark of the adapter disc (7)
l
Screw in the safety bolt (8) and tighten it
l
Remove existing type labels (when exchanging the emitter)
l
Unpack the type label set and letter it respectively. You can find
further information in chapter "Safety instructions - Labelling"
2
1
4
3
5
2
Fig. 17: Fasten type label
1
Type label - Source container
2
Type label - Source
3
Adhesive type label - Source
4
Adhesive label "Radioactive"
5
Clamping screws
A
Arrow mark
B
Straps must line up
l
Letter the adhesive type label (48 x 30 mm) and attach it to the
source holder
l
Letter the type label (107 x 17 mm) and fasten it to the neck of the
source container with two grooved pins
l
Attach the adhesive label "Radioactive" to the specified position of
the source container (if not already present)
l
Fasten the square adapters with flange (11) with the four clamping
screws (5). The arrow mark (A) of the source holder (6) is above
the OFF mark of the adapter disc (7)
l
Tighten the terminal screws (5)
Insert source holder
Attach or replace the
type labels
Mount squared adapters
and drive unit
Emitter loading and exchange • VEGASOURCE 31 35
19