l

Switch off compressed air and remove the feed lines. The source
container must now be in position OFF

l

Off position of the drive unit with padlock - secure switching
function (1)

l

Remove padlock (3)

l

Remove the seal (4) from the strap with side cutting pliers and pull
out the seal wire

l

Loosen fixing screws (2) (3 pieces)

l

Lift up the complete drive unit upward

l

Loosen fixing screws (5) (3 pieces)

l

Detach the squared adaters with flange (11) upward

12

13

8

14

13

Fig. 19: Emitter exchange - source holder VEGASOURCE version M, N
8

Source holder

12

Slot

13

Source capsule

14

Spring ring

l

Unscrew source holder (8) with spanner (size 13) and pull it out of
the source container

l

Only when exchanging the emitter: Insert the blade of a
screwdriver (4 mm) into the slot (12) on the source holder (8) and
press out the source capsule (13) or unscrew the source capsule
with a screwdriver (2 mm). Let the source capsule (13) fall into a
shielded vessel

Dismount drive unit and
squared adapter

Insert or exchange emit-
ter

22

Emitter loading and exchange • VEGASOURCE 31 35

2 Loading or emitter exchange

39155

-EN

-110210