VEGA VEGASOURCE 35 Emitter loading and exchange User Manual

Page 18

Advertising
background image

l

Switch off compressed air and remove the feed lines. The source
container must now be in position OFF

l

Off position of the drive unit with padlock - secure switching
function (1)

l

Remove padlock (3)

l

Remove the seal (4) from the strap with side cutting pliers and pull
out the seal wire

l

Loosen fixing screws - drive unit (2) (3 pcs, hexagon size 4) and
remove them. The screws have different lengths

l

Lift up the complete drive unit upward

l

Only loosen (4 pieces) the terminal screws (5) - do not unscrew
the screws

l

Detach the squared adaters with flange (11) upward

14

13

6

12

Fig. 16: Emitter exchange - source holder VEGASOURCE version K, L
6

Source holder

12

Safety bolt - Source capsule (optional)

13

Cover plate

14

Source capsule

A

Arrow mark

l

Loosen securing screw (8) with an Allen key size 4

l

Turn source holder (6) ¼ turn anticlockwise and pull it out of the
source container. The arrow mark (A) is above the OFF mark of
the adapter disc (7)

l

If it is present, remove the safety bolt (12) with Allen key size 2.5

l

Swivel the cover (13) in front on the thin cylinder of the source
holder to the side. To do this, lift the cover (13) slightly with a small
slotted screwdriver

l

Only when exchanging the emitter: Allow the source capsule (14)
to fall out downwards into a shielded vessel

l

Insert new source capsule (14) and close cover (13). For insertion
use a suitable gripping device or a manipulator

l

Screw in the safety bolt (12) (optional)

Dismount drive unit and
squared adapter

Insert or exchange emit-
ter

18

Emitter loading and exchange • VEGASOURCE 31 35

2 Loading or emitter exchange

39155

-EN

-110210