VEGA VEGASOURCE 35 Emitter loading and exchange User Manual

Page 9

Advertising
background image

l

Side cutters

l

Gripper tool or manipulator for source capsule

l

Possibly extension rod M8

l

Remove padlock (1)

l

Loosen locking screw (2) with a flat head screwdriver 1.5 x 10

l

Pull out fixing pin (3)

l

Remove safety catch (4)

l

Remove the seal between holding pin (5) and capstan screw (6)
with side cutting pliers and pull out the seal wire

l

Unscrew the capstan screw (5) with a flat headed screwdriver,
size 1.5 x 10

l

Insert the holding pin (5) and move with the stirrup over the pin

l

Bring the stirrup to the extraction position

Fig. 7: VEGASOURCE - version B in extraction position

l

If necessary, screw an extension rod (11) with thread M8 into the
socket of the source holder (10). With this extension you can
increase the distance to the emitter

l

Pull out the source holder (10)

l

If it is present, remove the safety bolt (7), Allen key size 2.5

l

Swivel the cover (8) in front on the thin cylinder of the source
holder (10) to the side. To do this, lift the cover (8) slightly with a
small slotted screwdriver

l

Only when exchanging the emitter: Allow the source capsule (9) to
fall out downwards into a shielded vessel

l

Insert new source capsule (9) and close cover (8). For insertion
use a suitable gripping device or a manipulator

l

Screw in the safety bolt (7) (optional)

l

Completely insert the source holder (10) into the source container
and turn it until the holding pin (5) pops out

l

If necessary, remove extension rod (11)

l

Continue turning the source holder (10) up to position OFF

Remove source holder

Insert or exchange emit-
ter

Insert source holder

Emitter loading and exchange • VEGASOURCE 31 35

9

2 Loading or emitter exchange

39155

-

EN

-110210

Advertising