Printronix Online Data Validator User Manual

Page 43

Advertising
background image

Scanning Beam Skew Adjustment

43

4.

Make sure the bar code is under the scanning beam.

5.

Press and hold the

RESET

button on the validator for 4 to 6

seconds until the READ light illuminates and the
CALIBRATION light begins flashing.

Figure 23. Calibrating the Validator

6.

Within a few seconds the CALIBRATION and READ light will
turn off. This indicates a successful calibration.

NOTE:

If the calibration light is still on or flashing, the calibration
has failed. In this case, refer to Table 2 on page 42 to
repeat the calibration procedure or to reset the validator
settings back to the last good calibration settings.

7.

Once you have a successful calibration, remove the calibration
bar code card.

183467a

CALIBRATION
Light

RESET Button

Deck Lock
Lever

Calibration Bar

Code Card