Xylem ProLab 3000 User Manual

Page 113

Advertising
background image

ProLab 3000

Ion concentration

113

ba75645e03

12/2012

6

In the Sample register, enter the values for the sample volume
(Sample volume) and the volume of the ISA/TISAB solution
(ISA/TISAB volume) as necessary.

7

Use [Dosing] to automatically add the adjusted volume of the
ISA/TISAB solution (ISA/TISAB volume).

The [Continue] button appears after the dosing.
or
If the burette control is inactive:

Add the adjusted volume of the ISA/TISAB solution (ISA/TISAB
volume
) manually.

8

Thoroughly rinse the electrode with deionized water.

9

Immerse the electrode in the sample.

10

Start the measurement with [Continue].
The measured value is checked for stability (Stability control).

11

Wait for the measurement with stability control to end.
or
Take over the calibration value with [OK].
The Standard register with the calculated measured value
pops up.

Advertising