3 general errors – Xylem ProLab 3000 User Manual

Page 176

Advertising
background image

What to do if...

ProLab 3000

176

ba75645e03

12/2012

Warning [TempErr]

Warning [ISEErr]

Burette does not dose

15.3

General errors

Meter does not react to

keystroke

Time is lost

Cause

Remedy

– Temperature difference

between measurement and cali-
bration greater than 2 K.

– Adjust the temperature of

the test sample

Cause

Remedy

– Electrode voltage outside cali-

brated range

– Recalibrate

Cause

Remedy

– Address of the burette incor-

rectly set

– Enter the address of the

burette on the meter

(see page 99)

– Enter the address of the

burette on the burette

(see operating manual of the
burette)

– RS232 interface of the burette

incorrectly set

– Configure the RS232 inter-

face of the burette

(see page 29 and operating
manual of the burette)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the <On/Off> and
<SC> key simultaneously.

Cause

Remedy

– Buffer batteries empty

– Exchange buffer batteries

(see page 171)