Xylem ProLab 3000 User Manual

Page 133

Advertising
background image

ProLab 3000

Ion concentration

133

ba75645e03

12/2012

Measuring

6

In the Sample register, enter the following values:
 the volume of the sample (Sample volume)
 the volume of the ISA/TISAB solution (ISA/TISAB volume)
 the volume of the blank value solution (Blank value volume)
 the concentration of the blank value solution (Blank value

conc.).

7

Use [Dosing] to automatically add the adjusted volume of the
ISA/TISAB solution (ISA/TISAB volume) and the blank value
volume (standard volume).
or
If the burette control is inactive:

Manually add the adjusted volume of the ISA/TISAB solution
(ISA/TISAB volume) and the blank value volume (standard vol-
ume).

8

Thoroughly rinse the electrode with deionized water.

Advertising