Ion concentration
ProLab 3000
126
ba75645e03
12/2012
12
In the Sample register, enter the values for the sample volume
(Sample volume) if necessary.
13
Use [Dosing] to automatically add to the standard solution the
adjusted volume of the sample (Sample volume).
or
If the burette control is inactive:
Add to the standard solution the adjusted volume of the sample
manually (Sample volume)
14
Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
15
Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.