Xylem ProLab 3000 User Manual
Page 114
Advertising
Ion concentration
ProLab 3000
114
ba75645e03
12/2012
12
In the Standard register, enter the values for Standard volume
and Standard concentration if necessary.
13
Use [Dosing] to automatically add to the sample the adjusted
volume of the standard solution (Standard volume)
or
If the burette control is inactive:
Add to the sample the adjusted volume of the standard solution
manually (Standard volume)
14
Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
15
Wait for the measurement with stability control to end.
or
Take over the calibration value with [OK].
The Calculation register with the calculated measured value
pops up.
Advertising