Xylem ProLab 3000 User Manual

Page 118

Advertising
background image

Ion concentration

ProLab 3000

118

ba75645e03

12/2012

10

Start the measurement with [Continue].
The measured value is checked for stability (Stability control).

11

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Standard register opens up.

12

Use [Dosing] to automatically add to the sample the adjusted
volume of the standard solution (Standard volume) (2 vol% of
sample solution).
or
If the burette control is inactive:

Add to the sample the adjusted volume of the standard solution
manually (Standard volume) (2 vol% of sample solution).

13

Start the measurement with [Continue].
The measured value is checked for stability (Stability control).

14

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.