ProLab 3000

Ion concentration

125

ba75645e03

12/2012

6

In the Standard register, enter the values for Standard vol-
umeStandard concentration
and ISA/TISAB volume if neces-
sary.

7

Use [Dosing] to automatically add the adjusted volume of the
ISA/TISAB solution (ISA/TISAB volume).
or
If the burette control is inactive:

Add the adjusted volume of the ISA/TISAB solution (ISA/TISAB
volume
) manually.

8

Thoroughly rinse the electrode with deionized water.

9

Immerse the electrode in the standard.

10

Start the measurement with [Continue].

The measured value is checked for stability (Stability control).

11

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Sample register opens up.