Xylem ProLab 3000 User Manual
Page 117
Advertising
ProLab 3000
Ion concentration
117
ba75645e03
12/2012
5
Thoroughly rinse the electrode with deionized water.
6
Immerse the electrode in the sample.
7
Start the measurement with [Continue].
After end of the measurement, the Standard pops up.
8
In the Standard register, enter the value for Standard concen-
tration if necessary.
9
Use [Dosing] to automatically add to the sample the adjusted
volume of the standard solution (Standard volume) (1 vol% of
sample solution).
or
If the burette control is inactive:
Add to the sample the adjusted volume of the standard solution
manually (Standard volume) (1 vol% of sample solution).