ProLab 3000

Ion concentration

129

ba75645e03

12/2012

6

In the Standard register, enter the values for Standard vol-
umeStandard concentration
and ISA/TISAB volume if neces-
sary.

7

Use [Dosing] to automatically add the adjusted volume of the
ISA/TISAB solution (ISA/TISAB volume).
or
If the burette control is inactive:

Add the adjusted volume of the ISA/TISAB solution (ISA/TISAB
volume
) manually.

8

Check the currently displayed ions.
If the ion of the standard (e.g. Cl (Sub) is displayed incorrectly:

Change the ion type of the standard solution [e.g. Cl (Sub)] in
the ION register.

9

Thoroughly rinse the electrode with deionized water.

10

Immerse the electrode in the sample.

11

Start the measurement with [Continue].

The measured value is checked for stability (Stability control).

12

Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Sample register opens up.