Xylem ProLab 3000 User Manual

Page 75

Advertising
background image

ProLab 3000

pH

75

ba75645e03

12/2012

Stability control

The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be termi-
nated at any time (accepting the current value).

Calibration record

When finishing a calibration, the new calibration values are first dis-
played as an informative message. Then you can decide whether you
want to take over these values of the new calibration or whether you
want to continue measuring with the old calibration data. After accept-
ing the new calibration values the calibration record is displayed.

Display of calibration

data and download to

interface

You can display the data of the last calibration (see page 142). Subse-
quently, you can transmit the displayed calibration data to the interface,
e. g. to a printer or PC, with the <PRINT> key.

Sample record:

Note
You can display the slope in the unit, mV/pH or % (see page 71).

Note
The calibration record is automatically transmitted to the interface
after calibrating.

Print date26.04.07 16 was selected:13
ProLab 3000 (06249876)
Current user: 1234567890

CALIBRATION pH
Calibration date20.04.07 10:14:03
A 161 1M-BNC-ID (A062498765)
User: Administrator

Calibration interval 7 days
AutoCal DIN 2-point TAuto SC
Buffer 1

6.86

Buffer 2

9.18

Voltage 1

7.2 mV 26.3 °C

Voltage 2

-171 mV 26.3 °C

Slope

-59.2 mV/pH

Zero point

-0.5 mV

Sensor

+++