Date – CEMB USA N300 User Manual

Page 22

Advertising
background image

4 - 2

Setup function

Note:

Although the instrument can operate correctly with any combination of sensors, we
recommend that sensors of the same type and model are connected to the two
channels.


For both possible types:

-

accelerometer

ACC

-

velocimeter

VEL

-

displacement

DIS


two different sensitivity levels can be set, identified respectively by “1” and “2”.
Therefore six different sensors can be defined:

-

ACC-1

accelerometer 1 (powered – IEPE type)

-

ACC-2

accelerometer 2 (powered – IEPE type)

-

VEL-1

velocimeter 1(not powered)

-

VEL-2

velocimeter 2(not powered)

-

DIS-1

displacement sensor 1 (not powered)

-

DIS-2

displacement sensor 2 (not powered)

making it possible to use the same types of sensors but with different sensitivity levels.


Typical sensitivity values for the various sensors are:

SENSOR TYPE

SENSITIVITY

TYPICAL VALUE

Accelerometer (ACC)

mV/g

100

Velocimeter (VEL)

mV/(mm/s)

21,2

Displacement (DIS)

mV/µm

8

Warning:

The sensitivity of some models may differ from the typical values shown; make
sure that the sensor is set with the correct value taken from its documentation.


Date

The date must be entered in the N300 instrument using two figures for the day, two for
the month and two for the year, in this order. To assist the operator, the dd/mm/yy
format is displayed below the current date.

New values can be entered following the same procedure as that described above for
the sensitivity of the sensors (see above).