Measurement results – CEMB USA N300 User Manual

Page 29

Advertising
background image

Vibrometer function

5 - 5

Measurement results


On the Vibrometer page the measurement results are displayed on the screen, successfully
combining the need for clarity and complete information.




Overall

measurement:

1. channel that the measurement displayed refers to
2. status of the channel displayed:

=

enabled

=

disabled

3. sensor connected to the channel
4. measurement taken (Overall)
5. number of averages
6. frequency band
7. vibration value, with unit and type
8. amplitude and frequency of the vibration’s highest component. The more

components the vibration is composed of, the more the overall measurement will be
greater than the amplitude of the dominating component. If the two values are very
similar, the vibration takes on an almost sinusoidal form.


The measurement can be started and stopped by pressing .

Note:

Values obtained this way can be used to assess the state of operation of the
instrument, using, for example, the tables and diagrams present in Appendix B of
this manual.






1



6

2 3 4

5



7


8

Advertising