Configuring the relays – Burkert Type 8025 User Manual

Page 54

Advertising
background image

54

operatingandcommissioning

Type 8025 / 8035

9.6.6. configuring the relays

This parameter makes it possible to set the switching thresholds and the operating behaviour, inverted or not, of each relay.

Enter values such as 1- ≤ 1+ and/or 2- ≤ 2+.

Both relays work in an hysteresis operating.

RELAy

FiLtER

2-= 0000

2+ = 0000

1+ = 0000

1+ = 0010

2+ = 0044

1-= 0000

iNv NO

2-= 0040

1-= 0008

iNv yES

ENTER

0......9

0......9

ENTER

ENTER

ENTER

0......9

ENTER

0......9

ENTER

iNv NO

iNv yES

0......9

0......9

Enter a flow rate value, associated to
the high threshold of relays 1, in the unit
chosen in the "UNIT" parameter.

Confirm.

ENTER

Choose the sense of operating,
inverted or not inverted.

Confirm

Choose the sense of operating,
inverted or not inverted.

Confirm.

Enter a flow rate value, associated to the low threshold of
relays 1, in the unit chosen in the "UNIT" parameter.

Confirm.

Enter a flow rate value, associated to
the low threshold of relays 2, in the unit
chosen in the "UNIT" parameter.

Confirm.

Enter a flow rate value, associated to
the high threshold of relays 2, in the unit
chosen in the "UNIT" parameter.

Confirm.

Fig. 61: Configuration of the relays threshold in hysteresis operating

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to

save the settings and go back to the Process level.

English

Advertising
This manual is related to the following products: