Maintenanceandtroubleshooting

Type 8025 / 8035

message
displayed

possible cause

recommended action

ERROR 7

Memory read error: the user settings
and the totalizer values are lost.

Press the ENTER key to acknowledge the error: the totalizers are
reset.

Send the device back to your supplier.

PWR FAIL

The supply voltage is too low.
• The voltage at the device terminals

is lower than 12 V.

• The impedance of the current

measurement loop is too high (see
chap. "6.5").

Adjust the supply voltage so that the voltage at the device ter-
minals is between 12 V and 36 V.

PU H LIM

The pulse value times the K factor of
the device is > 1000000.
The entered volume for a pulse is too
high.

Enter a lower volume / pulse. See chap. "9.6.5".

Check the K factor value. See chap. "9.6.3".

PU L LIM

The pulse value times the K factor of
the device is < 1.
The entered volume for a pulse is too
low.

Enter a higher volume / pulse. See chap. "9.6.5".

Check the K factor value. See chap "9.6.3".