Fig. 62: Hysteresis operating

9.6.7. configuring the filter of the measured flow rate

This parameter makes it possible to dampen the fluctuations:
• of the display;
• of the current output.
Ten filters are available.
The following table gives the response time for each filter (10% to 90%):

FILTER

Response time

FILTER

Response time

0

0,15 s

5

6 s

1

0,7 s

6

10 s

2

1,4 s

7

19 s

3

2,5 s

8

33 s

4

3,5 s

9

50 s

FiLtER

FiLtER 0

tOtAL

Choose the filter.

Confirm.

FiLtER 9

ENTER

0......9

Fig. 63: Diagram of the "FILTER" parameter of the Parameters menu

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to

save the settings and go back to the Process level.

9.6.8. resetting both totalizers

This parameter makes it possible to reset both totalizers.

tOtAL

RES NO

END

0......9

Choose to reset the two totalizers or not.

Confirm.

RES yES

ENTER

Fig. 64: Diagram of the "TOTAL" parameter of the Parameters menu