35

Adjustmentandcommissioning

A

Briefly press push button.

to calibrate automatically

Press and hold the push
button until...

to select the calibration of the
other parameter

B

Wait for 10 s.

to go back to the read mode

C

NO

Go back to Read mode
without confirming the
selection displayed.

Calibration in progress

Go back to Read mode.

.....

End of calibration: red LED
goes off?

YES

Zero point stored.

Red LED flashes briefly
twice: zero point has not
been calibrated.

Briefly press push button

Go back to Read mode
using the zero point of the
previous calibration.

Check that the calibration
instructions are respected
(see beginning chap. 9.7).

Restart calibration
of the zero flow
point.

...bargraph goes back to step 1, Fig.

28

Fig. 29: Calibration of the zero flow point, part 2