Adjustmentandcommissioning

Calibration in progress

Press and hold the push
button until...

to select the calibration of the
other parameter

B

A

Briefly press push button.

End of calibration: red LED
goes off?

Go back to Read mode

.....

NO

to calibrate automatically

Red LED flashes:
4 times: fluid velocity < 2 m/s
5 times: fluid velocity < 10 m/s
The full scale cannot be calibrated

Press push button for 2 s.

Go back to Read mode using the
full scale selected or previously
calibrated.

Full scale value stored

YES

Wait for 10 s.

to go back to the read mode

C

Go back to Read mode without con-
firming the selection displayed.

...bargraph goes back to step

1, Fig. 32

Fig. 33: Calibration of the full scale, part 2