Test Screens

There are three test screens used during a RATA: Error Calibration, Run, and Pre/Post
Calibration.

The Error Calibration screen (Figure 5) is used for the initial and final direct gas calibration of
the analyzers. The screen contains four list boxes for each analyzer, employing an indictator with
multiple lines of data, one of which can be highlighted. Only analyzers that have been configured
in the ‘Analog Input Configuration’ screen will be displayed in this test screen. The visible items
on the screen are adjusted in size to fill the available space in the Test Panel.

Figure 5


The four list boxes for each analyzer are used for the Zero, Low, Mid and High calibrations. The
analyzers that do not require a Low calibration will have that list box grayed-out. Prior to a
calibration the list box contains “Waiting For Cal”. In Manual mode the operator will control the
gas flow. In Automatic mode the computer controls the gas flow. While the specified cal gas is
flowing the list box will display “None Selected” at the top with 15 second average readings
adding just below “None Selected”. The newest reading is at the top of the list. The time of the
reading is displayed next to the 15 second average along with the letter “c”(cal error) if the
reading does not match the cal requirement. In Manual mode the operator will select the value to
be recorded for the calibration. In Automatic mode the computer will record the last reading after
a specified period of time. When the calibration value is selected the list box will be colored blue
if the reading meets the specified requirements, yellow if the reading is outside of the specifed
requirements.