Valve sequence order, Figure 22, Valve sequence order screen – Perma Pure RATAMATION V2.0 User Manual

Page 38: Ratamation v2.0 user manual perma pure llc

Advertising
background image

RATAmation v2.0 User Manual

Perma Pure LLC

Phone (732) 244-0010

www.permapure.com

Sept. 2012

36

Valve Sequence Order


‘Valve Sequence Order’ is a screen to adjust the order of use for the cal gas valves of the flow
drawer and the sequencing order of the calibration gases introduced during operation in
Automatic mode The order established in this screen is also used as a prompt during Manual
mode operation.


Open this screen from the Main Menu by selecting the ‘Configuration’ menu item and then
selecting ‘Valve Sequence Order’ from the submenu.

Figure 22

This screen will display all possible calibration gases including the divided gases in a scrollable
list. The list is shown in order from top to bottom with the cal gas valves as they will be used
during a calibration.


To adjust the valve order; highlight the valve to be moved and click the ‘Move Up’ or ‘Move
Down’ buttons. The highlighted valve will swap position with the valve directly above or below
respectively. Repeated clicks of the ‘Move Up’ or ‘Move Down’ buttons will continue to shift
the position of the highlighted valve.