When the screen is opened the current cylinder/analyzer information used in the program will be
used to populate the grid.
Each column will represent each of the analyzers. Each row will represent each of the calibration
cylinders. Each cell of the grid will contain a series of letters to represent the calibration usage.
The following letter code is used:
Z –Zero
Direct Calibrations
L – Low
M – Mid Range
H – High Range
Z – Zero
System Calibrations
S – Full Span
Each cell contains a combination the above letters or “-“marks. For example if cylinder 1 is used
for the High Range and Full Span then the cell would contain
“---H ;-S”
Selecting a cell of the grid will copy the data to fields where the info can be edited.
There are two sets of radio buttons in the edit area of the screen. Only 1 button of each set can be
chosen at a time, selecting an item will clear the other buttons of the set. The upper set specifies
the Error (Direct) Calibrations and consists of
N/A
Not used for calibration of this analyzer
Zero
Used for the zero calibration point
Low
Used for the low cal point, may be not available depending on the
analyzer gas measured
Mid
Used for the mid cal point
High
Used for the high cal point
The lower set specifies the Pre/Post (System) Calibrations and consists of
N/A
Not used for calibration of this analyzer
Zero
Used for the zero calibration point
Span
Used for the span cal point
The range and type input of the corresponding analyzer and the components of the corresponding
cal gas for the selected cell are displayed next to the radio buttons.