Calibration Gas Configuration


“Calibration Gas Configuration” is a screen to define the gases used by the flow drawer for
calibration. (Figure 19)

Open this screen from the Main Menu by selecting the ‘Configuration’ menu item and then
selecting ‘Configure Cal Gas’ from the submenu.

Figure 19

This screen will allow configuration of up to 14 calibration gases labeled as CG 1 though 14
(abbreviation for calibration gas) and 28 divided gases labeled DG 1 through 28.

CG 1 through 14 are connected to the flow drawer at cal gas valves in the rear of the unit. DG 1
through 28 can be connected to any of the 14 cal gas valves. The valve used for the divided gas
is set in the Program Set Up screen. The software only allows for the passage of the divided gas.
The gas divider itself must be controlled externally from this software.

The configurations for all cal gases are displayed in a spreadsheet type grid.

The screen is initially displayed with the stored setup information used by the program in the
main grid and with CG 1 selected.