54

2. If you are connecting a Model 4P-4300 Flow Control Drawer to a dedicated computer, the

TCP/IP settings must be set explicitly. This must be done for the settings on the computer AND
the flow control drawer.

a. Turn off Windows Firewall until communications have been established.
b. To set the TCP/IP settings to automatic, go to Start, Select Control Panel, select

Network Sharing Center, in the upper left corner select Change Adapter Settings and
a window will open displaying your network connections. Right click on local area
connection and select Properties. A dialogue box will open. Select Internet Protocol
Version 4 (TCP /IPv4)
and click on the Properties button. Another dialogue box
window will open.

c. There are two selection points. Select Use the following IP address: The DNS server

address can be left blank.

i. Enter the IP Address as

192.168.135.10

ii. Enter the Subnet Mask as

255.255.255.0

iii. Enter the Default Gateway as 192.168.135.9

d. Click OK and close out of all windows.

e. On the RATAmation software disc there is a program called SetupNE3.exe. Click to

install and follow directions to complete installation.