Teledyne 9060L - Zirconium oxide flue gas oxygen analyzer User Manual

Page 19

Advertising
background image

Percent Oxygen Analyzer

Operational Theory

Teledyne

Analytical

Instruments

19

require a heater. When the outside and inside of the sensor are exposed
to different oxygen partial pressures, an EMF (E) is developed which
obeys the Nernst equation:

⎟⎟

⎜⎜

=

Outside

)

(PO

Inside

)

(PO

log

4F

RT

ts)

E(millivol

2

2

e


where T is the temperature (K) at the disc (>650°C (1200°F)), R is the
gas constant. F is the Faraday constant and (PO

2

) Inside and (PO

2

)

Outside are the oxygen partial pressures at the inner and outer electrodes
respectively with the higher oxygen partial pressure being positive.

If dry air at atmospheric pressure, (21% oxygen) is used as a

reference at the inner electrode, the following equations are obtained:

⎟⎟

⎜⎜

×

=

Outside

)

PO

(

21

.

0

log

T

10

154

.

2

ts)

E(millivol

2

e

2

Transposing this equation:

=

T

46.421E

-

EXP

0.21

(ATM)

Outside

)

%O

(

2

The 9060L transmitter solves this equation which is valid above

650°C (1200°F). Either the probe heater or the process maintains the
sensor temperature at this level.

2.2.3 Heater

CAUTION: DANGEROUS VOLTAGE (120 VAC) EXISTS WITHIN

THE PROBE OR HEATER AND PRESENTS AN
ELECTRICAL SHOCK HAZARD TO MAINTENANCE
PERSONNEL. ALWAYS DISCONNECT THE PROBE
FROM THE ANALYZER BEFORE WORKING WITH
THE PROBE, SENSOR, OR HEATER.

THE EARTH CONNECTION (GREEN WIRE) FROM

Advertising