Teledyne 9060L - Zirconium oxide flue gas oxygen analyzer User Manual

Page 26

Advertising
background image

Installation Model

9060L

Teledyne Analytical Instruments

26

Figure 3-2: Model 9060L Rear panel

3.4.2 Analog Outputs

There are three DC output signal connectors with screw terminals

on the panel. There are two wires per output with the polarity noted. See
Figure 3-3. The outputs are:
0–10V % Range:

Voltage rises with increasing oxygen
concentration, from 0V at 0 percent oxygen to
10V at full scale percent oxygen. (Full scale =
100% of programmed range.)

0–10V Range ID:

03.33V = Low Range, 06.66V = High Range,
10V = Air Cal Range.

4–20 mA % Range: Current increases with increasing oxygen

concentration, from 4 mA at 0 percent oxygen to
20 mA at full scale percent oxygen. (Full scale =
100% of programmed range.)

3.4.3 Alarm Relays

The three alarm-circuit connectors are screw terminals for making

connections to internal alarm relay contacts. There is one set of contacts
for each type of alarm. Contacts are Form C, with normally open and
normally closed contact connections capable of switching up to 1.0
ampere at 125 VAC into a resistive load (2A for 30 VDC).