4.6.2 Air Calibration

To perform air calibration, make sure you have air flowing through the

same inlet you use for span calibration at a flow of 1- 5 liters per minute. Wait
for readings to stabilize.

Press both the SET ALARM 1 and SET ALARM 2 simultaneously. The

LED display will show “OFSt” for a few seconds. After that the instrument
should read near 20.9%.

The sensor output is expected to be near zero millivolts, if the offset of

the sensor is too high, the analyzer will not calibrate.

4.7 Supplementary Information

If, during the span procedure, you pressed the SPAN button by

mistake, you must wait five minutes for the analyzer to resume analysis
or you can press the UP button and then the DOWN button. (Pressing
the UP and DOWN buttons causes the analyzer to time-out in five
seconds instead of five minutes).

If during the span procedure, you press the RANGE or ALARM

buttons:

- either the range or alarm routine will be activated.

- any changes to span will be rejected.

- the 60 second alarm delay will not occur, i.e., the alarms

will be responsive immediately.

CAUTION: TI/AI CONSIDERS THE ACTION OF PRESSING THE

ALARM OR RANGE BUTTONS TO BE AT YOUR (THE
USER’S) DISCRETION AND NOT UNDER
GUARANTEE OF ALARM PROTECTION.

4.8 Probe Temperature

The probe temperature can be displayed momentarily and adjusted

if needed. While the instrument is in normal mode, displaying oxygen
concentration, press the UP and DOWN arrow keys simultaneously. The
red LED display will show the probe temperature in degrees centigrade
for five seconds.