Teledyne 9060L - Zirconium oxide flue gas oxygen analyzer User Manual

Page 44

Advertising
background image

Appendix Model

9060L

Teledyne Analytical Instruments

44

A.2 Probe Specifications

Application:

Combustion flue below 700°C (1290°F),
above 700°C (1290°F) 900°C (1650°F) — no
contaminants.
With contaminants see notes 1 &2

Temperature

Range:

0-900°C

800-1400°C 800-1200°C

(32-1650°F) (1470-2550°F) (1470-2190°F)

Length:

250-1000 mm 500-1160 mm 457-1219 mm
(10” to 40”) (20” to 46”) (18” to 48”)

Process

Connection:

1 1/2" BSP 3/4" BSP 1" BSP or equivalent
NPT

Electrical

Connection:

Weatherproof plug-in connector or
optional screw terminals. Plug connector
supplied with the cable. Ex d heads have
screw terminals.

Cable: Order a specific length with the analyzer

except for hazardous installations where
the cable is supplied by the customer.

Heater: Yes

Flue Gas Thermocouple: K

Response

Time:

Typically < 4 secs.

Head Temperature: 125°C (250°F)

Reference

Gas: Ambient air 50 to 500 cc/min (6 to 60

scfm). Pump supplied with analyzer

Ref Air Connection: 1/4" NPT integral air line in probe cable.

Barbed fitting to 3/16" ID PVC tube.

Filter:

Removable sintered stainless steel
particulate filter, 15 micron.

Calibration Gas Connection: 1/8" NPT female 1/8" NPT female 1/8"

NPT female

Weight

:

0.6 kg (1.32 lbs.)