4 setting the alarm setpoints – Teledyne 9060L - Zirconium oxide flue gas oxygen analyzer User Manual

Page 33

Advertising
background image

Percent Oxygen Analyzer

Operation

Teledyne

Analytical

Instruments

33

range full scale value, and it switches back to the LO range when the
oxygen concentration reaches 85% of the LO range full scale value

Note: The HI Range setpoint MUST be set at a higher

concentration percentage than the LO Range setpoint.

4.3.1 HI Range

Setting the HI Range full scale value defines the LEAST sensitive

analysis range to be used. To set the HI Range:

1. Press the SET HI RANGE function button once.

2. Immediately (within 5 seconds) press either the

Δ or ∇ button

to raise or lower the displayed value, as required, until the
display reads the desired full scale percent concentration.

4.3.2 LO Range

Setting the LO Range full scale value defines the MOST sensitive

range to be used. To set the LO Range:

1. Press the SET LO RANGE function button once.

2. Immediately (within 5 seconds) press either the

Δ or ∇ button

to raise or lower the displayed value, as required, until the
display reads the desired full scale percent concentration.

4.4 Setting the Alarm Setpoints

The alarm setpoints can be adjusted over the full range of the

analyzer (0-100% oxygen content). They are set as a percent of oxygen
content, so that an alarm set to indicate 9.6 on the display will activate at
9.6% O

2

on any O

2

range.

4.4.1 Set Alarm 1 (HI alarm)

Alarm 1 is preset as a high alarm. To configure this alarm to your

preferences:

1. Press the SET ALARM 1 function button once.
2. To change the setting at which the alarm will be actuated, press

the SET ALARM 1 function button once more. The alarm
setpoint will flash on the LED display. Press either the Up or
Down keys to raise or lower the displayed value, as required,