Teledyne 9060L - Zirconium oxide flue gas oxygen analyzer User Manual

Page 47

Advertising
background image

Percent Oxygen Analyzer

Appendix

Teledyne

Analytical

Instruments

47

11.0

13.78

10.5

14.78

10.0

15.82

9.5 16.92

9.0

18.08

8.5

19.30

8.0

20.60

7.5

21.98

7.0

23.45

6.5

25.04

6.0

26.75

5.5

28.61

5.0

30.65

4.5

32.90

4.0

35.42

3.5

38.28

3.0

41.58

2.5

45.48

2.0

50.25

1.5

56.41

1.0

65.08

0.5

79.91

0.2

99.51

'K’ TC mV 29.212 at 720 °C (1320°F)
These tables are based on the Nernst equation:

Sensor e.m.f. = 0.02154 x T x 1n ( 20.95 / % oxygen),

where T = ° K (° C + 273), e.m.f. is in mV’s