Teledyne 9060L - Zirconium oxide flue gas oxygen analyzer User Manual
Page 47
Advertising
Percent Oxygen Analyzer
Appendix
Teledyne
Analytical
Instruments
47
11.0
13.78
10.5
14.78
10.0
15.82
9.5 16.92
9.0
18.08
8.5
19.30
8.0
20.60
7.5
21.98
7.0
23.45
6.5
25.04
6.0
26.75
5.5
28.61
5.0
30.65
4.5
32.90
4.0
35.42
3.5
38.28
3.0
41.58
2.5
45.48
2.0
50.25
1.5
56.41
1.0
65.08
0.5
79.91
0.2
99.51
'K’ TC mV 29.212 at 720 °C (1320°F)
These tables are based on the Nernst equation:
Sensor e.m.f. = 0.02154 x T x 1n ( 20.95 / % oxygen),
where T = ° K (° C + 273), e.m.f. is in mV’s
Advertising