Teledyne 9060L - Zirconium oxide flue gas oxygen analyzer User Manual

Page 32

Advertising
background image

Operation Model

9060L

Teledyne Analytical Instruments

32

4.2 Using the Function and Data Entry Buttons

When no buttons on the analyzer are being pressed, the instrument

is in the analyze mode. It is monitoring the percent of oxygen in the
sample gas that is flowing through the remote probe.

When one of the function buttons is being pressed, the analyzer

enters the setup mode or the calibration mode.

Setup Function Buttons:

There are 4 setup function buttons on the analyze:

• SET ALARM 1
• SET ALARM 2
• SET HI RANGE
• SET LO RANGE

Calibration Button:

The calibration mode button is:

• SPAN

Data Entry buttons:

The data entry buttons (

Δ and ∇) increment the values displayed on

the PERCENT OXYGEN meter while one of the function buttons is
being held down.

• Δ : Increments the displayed value upwards.
• ∇ : Increments the displayed value downwards.

Any of the functions can be selected at any time by holding down

the appropriate button.

Each function will be described in the following sections. Although

the operator can use any function at any time, the order chosen in this
manual is appropriate for an initial setup.

4.3 Setting the Analysis Ranges

The two user definable analysis ranges are both capable of being

adjusted for from 0-1 to 0-100% oxygen concentration. Whatever values
are selected, the analyzer automatically switches from the LO range to
the HI range when the oxygen concentration reaches 100% of the LO