Indoor Setup

Overview

This section contains instructions for setting up a portable
or indoor system using the W1314A receiver. Always charge
all batteries before beginning a survey.

The indoor hardware includes:

A standard backpack

Two batteries (one battery is fitted with a power booster

module)

One Agilent W1314A receiver

One timing module (1PP2S output)

Cables and connections

The timing module provides a 1PP2S signal for the receiver
measurements.

Setting up the timing module

Before the timing module can be used, the 1PP2S output
needs to be calibrated against GPS timing signals. This can
take between 4 and 8 hours depending on GPS signal
strength.

Follow this procedure to calibrate your timing module:

1 Install the GPSMon

1

software on any compatible laptop or

PC. Refer to the manufacturer’s installation instructions.

2 Connect the timing module to your laptop using the

RS232 port. Refer to

Figure 22

on page 48.

1 The GPSMon software and documentation can be found in the following

directory and folder:
C:\Program Files\Agilent Technologies\E6474A-X\Drivers\GPSMon
Double click on GPSMon.msi to start the installation process.