<7. Startup>

7-13

IM 11M12A01-02E

7.11.2 Checking Calibration Contact Output

The calibration contacts are used for solenoid valve drive signals for the ZR40H Automatic

Calibration Unit. When using the ZR40H Automatic Calibration Unit, use the calibration contact

output to check that the wiring connections have been properly completed and check equipment

operation.
(1) Referring to Section 7.11.1, display the contact check display.
(2) Select the “Calibration contacts” to display the panel display as Figure 7.15.3 shows.
(3) Open the “Zero gas contact” and the “Span gas contact”. This will help check the automatic

calibration unit and wiring connections.

Zero gas contact : Open

Span gas contact : Open

Enter

Calibration contacts

F7.15.3E.ai

Figure 7.15.3 Calibration Contact Check Display

NOTE

“Open” and “Closed” displayed on the Calibration contacts display indicate actions of drive

contacts and are opposite to the valve open and close actions. If “Open” is displayed on the

Calibration contacts display, no calibration gas flows. If “Closed” is displayed on that display,

calibration gas flows.

7.11.3 Checking Input Contacts

(1) Referring to Section 7.11.1, display the contact check display.
(2) Display the “Input contacts” check display as Figure 7.15.4 shows. The “Open” or “Closed”

input contact in the display shows the current contact input terminal status and the display

changes according to the contact status. Using this enables you to check that the wiring

connections have been properly completed.

Input contact 1 : Open
Input contact 2 : Open

Enter

Input contacts

F7.15.4E.ai

Figure 7.15.4 Input Contact Check Display