3 installation of the detector, Installation of the detector -2, K9471uc), probe protector zo21r -2 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 48: Caution

Advertising
background image

<3. Installation>

3-2

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

(1) Do not mount the probe with the tip higher than the probe base.
(2) If the probe length is 2.5 meters or more, the detector should be mounted vertically

(no more than a 5° tilt).

(3) The detector probe should be mounted at right angles to the sample gas flow or the probe

tip should point downstream.

100 mm

100 mm

F3-1E.ai

(vertical)

Bounds of the probe

insertion hole location

Flange matches

the detector size

(horizontal)

Note

Type

Outside diameter of detector

Standard

50.8 mm in diameter (Note)

With dust filter

51 mm in diameter (Note)

With probe protector 60.5 mm in diameter (Note)

Four-hole flange Eight-hole flange

*1

*1

(Note)

When using the detector with pressure compensation, ensure that the flange gasket does not block the reference gas outlet on

the detector flange. If the flange gasket blocks the outlet, the detector cannot perform pressure compensation.

Where necessary, make a notch in the flange gasket. Confirm the outside dimensions of the detector in Section 2.2 before

installation.

Figure 3.1 illustrates an example of the probe insertion hole.

3.1.3

Installation of the Detector

CAUTION

• The cell (sensor) at the tip of the detector is made of ceramic (zirconia). Do not drop the

detector, as impact will damage it.

• A gasket should be used between the flanges to prevent gas leakage. The gasket material

should be heatproof and corrosion-proof, suited to the characteristics of the sample gas.

The following should be taken into consideration when mounting the general use detector:
<General use detector>
(1) Make sure that the cell mounting screws (four bolts) at the tip of the detector are not loose.
If a dust filter (see Section 2.7.1) is used, make sure it is properly attached to the detector.

Refer to Section 3.1.4 for installation of the dust filter.
(2) Where the detector is mounted horizontally, the calibration gas inlet and the reference gas

inlet should face downward.

3.1.4

Installation of the Dust Filter (K9471UA), Dust Guard

Protector (K9471UC), Probe Protector ZO21R