2 wiring for detector output, 1 cable specifications, Wiring for detector output -5 5.2.1 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 77: Cable specifications -5, Caution

Advertising
background image

<5. Wiring>

5-5

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

5.2

Wiring for Detector Output

This wiring enables the converter to receive cell output from the detector, output from a

thermocouple and a reference junction compensation signal. Install wires that allow for 10 Ω of

loop resistance or less. Keep detector wiring away from power wiring.

CELL(+)

CELL(+)

CELL(-)

CELL(-)

TC(+)

TC(+)

TC(-)

TC(-)

CJ(+)

CJ(+)

CJ(-)

CJ(-)

FG

Shielded cables

Shielded cables

ZR22G

Detector

ZR402G

Converter

CELL(+)

CELL(+)

CELL(-)

CELL(-)

TC(+)

TC(+)

TC(-)

TC(-)

CJ(+)

CJ(+)

CJ(-)

CJ(-)

FG

(2) Ambient temperature of the detector: exceeding 80 °C

Heat-resistant wiring

Terminal box

(1) Ambient temperature of the detector: 80 °C or less

F5-5E.ai

ZR22G

Detector

ZR402G

Converter

Figure 5.4 Wiring for detector output

CAUTION

If shielded cables cannot be used between the detector and the terminal box, for example,

when heat-resistant wiring is used, locate the detector and the terminal box as close together as

possible.

5.2.1

Cable Specifications

Basically, PVC insulated PVC sheathed cable (6-core) is used for this wiring. When the ambient

temperature of the detector exceeds 80 °C, install a terminal box, and connect with the detector

using six-piece 600 V silicon rubber insulated glass braided wires.

Advertising
This manual is related to the following products: